ef:

The ef command will put the instrument in a "Fixed final energy" configuration.

The syntax of the ef command is:

ef value

where value is the stored energy value.  Note that this command does not drive the analyzer to this value of ef - to do so, type:

drive ef value



The ef command can also be accessed from the GUI by selecting the Experiment tab from the Setup top-level tab.  To change to a fixed final energy configuration and/or the stored value of Ef, the green button at the bottom of the tab must be changed from "Monitoring ... Click to allow changes" to "Changing ... click to monitor".  Select "Fixed Final Energy (meV)" from the pull-down menu, change the value for "Fixed Energy (meV)" and commit the changes by clicking on the blue "Update Parameters" button.

GUI interface to ef