plusminus:

The plusminus command is used to specify the direction of scattering for the monochromator, sample, and analyzer.

The syntax for the plusminus command is:

plusminus +/-1 +/-1 +/-1   

OR   

plusminus +/- +/- +/-

The monochromator scattering direction is fixed for a given instrument and cannot be changed using this command.

NOTE: All angles are positive in the trigonometric (i.e. counterclockwise) sense.  The standard W configuration is either plusminus + - + OR plusminus - + - depending on the sign of the monochromator scattering direction.



The plusminus command can also be accessed from the GUI by selecting the Experiment tab from the Setup top-level tab.  To change the scattering directions for the instrument, the green button at the bottom of the tab must be changed from "Monitoring ... Click to allow changes" to "Changing ... click to monitor".  Select either Plus or Minus from the blue pull-down menus and commit the changes by clicking on the blue "Update Parameters" button.   NOTE that the MONOCHROMATOR pull-down menu will not allow you to change the initial scattering direction as this is fixed by physical constraints on the instrument.

GUI interface to plusminus